Autoclave, Class N, Horizontal Type
SKU: STH-ND Series
- Horizontal type, Jacket structure
- Chamber volume: 150L,200L, 300L, 400L
- Working temperature: 105-138℃
- Working pressure: 0.245 Mpa
- Can set multiple sterilization programs
- Multiple safety protection
- Horizontal circular inner chamber and jacket structure, safety, heat preservation, and drying function;
- Disc type embedded gear and Electric lock door structure, safety and reliable;
- Door seals are pneumatic
- Equipped with standard equipment verification interface
- Pipeline system for the optimal design, SUS 304 pipeline, clamp structure, quick assembly
- Control system: HMI man-machine interface, touch screen + microcomputer;
- FO value and temperature time double guarantee sterilization effect, with perfect sterilization records;
- With program selection, parameter setting, equipment operation, report processing and traceability Functions;
- Multiple safety protection(over-temperature, over-pressure protect)
- Built-in printer
- USB function
Specifications
Model Volume(L) Overall Dimensions
(L×W×H)mmChamber Size
(Ø×L)mmSteam Consumption
(KG)Water Consumption
(KG)Power
(KW)Net Weigh
(KG)STH-N150D 150 1290×690×1680 Ø472×880 ≤12 ≤190 ≤2+⑫ 350 STH-N200D 200 1560×690×1680 Ø472×1150 ≤13 ≤190 ≤2+⑫ 370 STH-N300D 300 1430×850×1760 Ø620×995 ≤18 ≤200 ≤2+⑮ 450 STH-N400D 400 1755×850×1760 Ø620×1320 ≤22 ≤200 ≤2+⑮ 510 Main Technical Parameter
Design pressure 0~0.3Mpa Design temperature 144℃ Working pressure 0.245Mpa Working Temperature 105-138℃ Temp accuracy ±1℃ Temperature Equilibrium <+2.0℃ Water Pressure 0.1-0.2Mpa Compressed air pressure 0.4~0.7Mpa Steam Pressure 0.3-0.7Mpa Power voltage Three-phase five-wire AC 380V 50HZ Note: The machine needs external compressed air for door sealing and the air consumption is not large. If the end users don’t have a ready air source, they can choose to use an air compressor of about 40L.